FULLY INTEGRATED R-3000/R-3000HR System includes:
- Laser (785nm/532nm) (Details below)
- Spectrometer
- Linear array CCD detector
- Fiber optic sampling probe
- Sample compartment with holder for vials and cells
- Sampling caps for liquids and solids, standoff and immersion measurements
- Verification cap, for one-touch verification and calibration
- Safety goggles
- RamanDesktop software with library building capability
- Laptop computer included
- 1-year warranty
|
Laser specifications
-
Excitation wavelength: 785 nm or 532 nm
-
Output power: 250mW (785nm) and 25mW (532nm)
-
Software controlled “power selection” switch
-
Laser linewidth < 0.5nm
-
Thermoelectrically cooled and grating stabilized
-
Reproducibility better than 1 cm-1
Spectrometer specifications
-
Spectral Resolution: Better than 10 cm-1 (R-3000) and better than 6 cm-1 (R-3000HR)
-
Dynamic Range: 200-2700 cm-1 (785nm) and 200-4000 cm-1 (532nm)
-
X-axis stability: Better than 1 cm-1.
-
Y-axis stability: Better than 5%
Probe and Sampling
-
Fiber optic probe with software controlled safety shutter, for safety of personnel
-
Interchangeable
measurement caps for direct-contact measurement of solids, standoff
measurement through vials and cells or immersion measurements into
liquids.
Other specifications
Software Features
Runs on RSIBase software.
Warranty
All
items manufactured by Raman Systems Inc. (end-user products, OEM
modules and complete systems) are warranteed to be free from defects in
materials and workmanship. This warranty is valid for a period of 12
months from the date of delivery of the order.
The warranty does
not include the computer or products manufactured by third parties. The
warranty on the computer is transferred directly to the end-user and
hence end-users are requested to contact the computer manufacturer
directly in connection with any service related issues.
|